Mass Spectrometry Services at Venture Center are provided by the The Center for Applications of Mass Spectrometry (CAMS); a facility developed and managed by the BioIncubator at Venture Center, NCL Innovation Park, Pune, India as a resource center for fostering collaboration between researchers and industrial/ entrepreneurial organizations leveraging mass spectrometry.
CAMS has a specific focus and emphasis on technology advancement, industrial applications and new venture creation (especially services leveraging mass spectroscopy to create value and tools enhancing the applicability of mass spectrometry). The Center aims to foster collaboration between researchers and industrial/ entrepreneurial companies.

Agilent 6550 UHD Accurate Mass QTOF MS

- Mass Analyzer: Quadrupole Time Of Flight (QTOF)
- Ionisation Sources: ESI, APCI, APPI , ASAP
- Detector: Electron multiplier
- Mass accuracy within 1 ppm; m/z (z=1) less than 500
- MS/MS accuracy within 2 ppm
- S/N ratio of 450:1
- Resolution of 42,000 at m/z 2722
- High Femtogram level sensitivity, I pg at m/z 609
- Dynamic range of up to five orders
- Ability to scan multiple precursor ions ~50
- Rapid polarity switching of 1.5 sec
- Scan rates of 50 Hz
- Positive and negative ion modes

Agilent 6540 UHD Accurate Mass QTOF MS

- Mass Analyzer: Quadrupole Time Of Flight (QTOF)
- Ionisation Sources: ESI, APCI, APPI , ASAP
- Detector: Electron multiplier
- Mass accuracy within 1 ppm; m/z (z=1) less than 500
- MS/MS accuracy within 2 ppm
- S/N ratio of 450:1
- Resolution of 42,000 at m/z 2722
- High Femtogram level sensitivity, I pg at m/z 609
- Dynamic range of up to five orders
- Ability to scan multiple precursor ions ~50
- Rapid polarity switching of 1.5 sec
- Scan rates of 50 Hz
- Positive and negative ion modes

Agilent 1260 Infinity Nanoflow LC

- G4225A 1260 HiP Degasser
- G1376A 1260 Cap Pump
-
Flow rate between 0.01 to 20.0 μl per min
-
Pressure range 20 to 400 bar (5880 psi) system pressure
-
Recommended pH range 1.0 – 8.5, solvents with pH < 2.3 should not contain acids which attack stainless steel. Upper pH range is limited by fused silica capillaries.
- G4446A 1260 Nano Flow pump
-
Flow rate between 0.01 to 4.0 μl per min
-
Pressure range 20 – 400 bar (5880 psi) system pressure
-
Flow Precision :< 0.2 % RSD.
- G1330B 1290 Thermostat
-
Temperature range Settable from 4 °C to 40 °C in 1 ° increments
- G1377A 1260 HiP MicroALS
-
Sample capacity 2 mL vials in 2 x 54 vial plate
-
Injection range 0.01 – 8 µL in 0.01 µL increments
-
Pressure range Up to 400 bar (5880 psi)

Agilent 1260 Infinity Binary LC system

-
Flow range 0.05 – 5.0 mL/min
-
Flow precision< 0.07% RSD
-
Pressure operating range 600 bar
-
Gradient formation High-pressure binary mixing
-
Autosampler capacity 100 vials Injection volume of 0.1‐100 µl in 0.1µl increments
-
pH range 1.0‐12.5

Agilent CHIP cube interface

-
Microfluidic chip-based technology specifically designed for nanospray LC/MS.
-
Ambient operating temperature 5 – 40 °C
-
orthogonal dual electrode nanospray ion source
-
Integrated camera and light source directly in spray chamber.
-
Video output to external monochrome monitor for spray visualization

Agilent GC-MSD

7890 GC
-
Split / Splitless capillary inlet: Split ratio of up to 7,500:1
-
Column oven operating temperature range + 4 to 450°C
-
Temperature set point resolution 0.1°C
-
Supports 20 oven ramps
-
Maximum achievable temperature ramp of 120°C/min
5977B MSD
-
EI source
-
Ion source temperature 150–350 °C
-
Quadrupole temperature 106–200 °C
-
Mass Filter Heated monolithic hyperbolic quadrupole
-
Mass range 1.6–1,050 u
-
Scan speed up to 20,000
-
Mass accuracy 1-μL injection of a 100 pg/μL OFN1 standard at m/z 271.987 ± 0.0052
-
Mass axis stability better than 0.10 u/48 hours
-
Detector Triple-Axis Detector with high energy dynode and long life electron multiplier
-
Software GC/MSD MassHunter Acquisition
-
Spectral Library NIST 2017

Agilent 7697A Headspace Sampler

-
Sample handling 12 vial capacity
-
Single position, solid aluminum vial oven
-
Chromatographic performance <1.5% RSD
-
Oven Off, 35 °C to 210 °C
-
Valve and Loop Off, 35 °C to 210 °C
-
Transfer Line Off, 35 °C to 250 °C
You can learn more about our services and pricing here